Replacement Circuit Board for Tube Holder S

Replacement Circuit Board for Tube Holder S, 4008573 [U99999-611], Electron Tubes S
Replacement Circuit Board for Tube Holder S, 4008573 [U99999-611], Electron Tubes S
97.17
In stock
The quality of the electron beam in electron defraction tube S (1013889) is affected by the resistance between sockets C5 (cathode) and F4 (heating filament) on the tube. For optimum results, the resistance needs to be 390 kΩ. Tube holder S (1014525) is accordingly designed such that this resistance is present.
Older designs feature a much smaller resistance and need to be modified in order to work with the new S-series electron deflection tube 1013889).
Tube holders affected:       U18500, U185001, 1000610

Required

 
557.19 €
 
 
 
Item No.: 4008573 [U99999-611]
Weight 0.05 kg
Brand 3B Scientific Teltron
  • Secure online payment with SSL
  • Expert advice
  • Easy returns & exchanges
  • International shipping available
Service hotline: +49-40-73966-223
Add to cart

Customers who bought this item also bought these items:
Electron Diffraction Tube S, 1013889 [U185711], Electron Tubes S
1,254.60 €
by 3B Scientific Teltron
Helmholtz Coils S, 1000611 [U185051], Experiments with Electron Tubes
523.98 €
by 3B Scientific Teltron
Thomson Tube S, 1000617 [U18555], Electron Tubes S
1,151.28 €
by 3B Scientific Teltron
Deflection Tube D, 1000651 [U19155], Electron Tubes D
1,161.12 €
by 3B Scientific Teltron
Tube Holder S, 1014525 [U185002], Electron Tubes S
557.19 €
by 3B Scientific Teltron